The use of electrothermal vaporizer coupled to the inductively coupled plasma mass spectrometry for the determination of arsenic,selenium and transition metals in biological samples treated with formic acid
Authors:Luciano Tormen  Raul A. Gil  Vera L.A. Frescura  Luis Dante Martinez  Adilson J. Curtius
Affiliation:1. Departamento de Química, Universidade Federal de Santa Catarina, Depto. Química, Campus Trindade, 88040-900 Florianópolis, SC, Brazil;2. Universidade Federal da Fronteira Sul – UFFS, Campus Laranjeiras do Sul, 85303-775 Laranjeiras do Sul, PR, Brazil;3. Instituto de Quimica de San Luis (UNSL-CONICET), Chacabuco y Pedernera, D5700BWQ San Luis, Argentina
Abstract:A fast method for the determination of As, Co, Cu, Fe, Mn, Ni, Se and V in biological samples by ETV-ICP-MS, after a simple sample treatment with formic acid, is proposed. Approximately 75 mg of each sample is mixed with 5 mL of formic acid, kept at 90 °C for 1 h and then diluted with nitric acid aqueous solution to a 5% (v/v) formic acid and 1% (v/v) nitric acid final concentrations. A palladium solution was used as a chemical modifier. The instrumental conditions, such as carrier gas flow rate, RF power, pyrolysis and vaporization temperatures and argon internal flow rate during vaporization were optimized. The formic acid causes a slight decrease of the analytes signal intensities, but does not increase the signal of the mainly polyatomic ions (14N35Cl+, 14N12C+, 40Ar12C+, 13C37Cl+, 40Ar36Ar+, 40Ar35Cl+, 35Cl16O+, 40Ar18O+) that affect the analytes signals. The effect of charge transfer reactions, that could increase the ionization efficiency of some elements with high ionization potentials was not observed due to the elimination of most of the organic compounds during the pyrolysis step. External calibration with aqueous standard solutions containing 5% (v/v) formic acid allows the simultaneous determination of all analytes with high accuracy. The detection limits in the samples were between 0.01 (Co) and 850 μg kg−1 (Fe and Se) and the precision expressed by the relative standard deviations (RSD) were between 0.1% (Mn) and 10% (Ni). Accuracy was validated by the analysis of four certified reference biological materials of animal tissues (lobster hepatopancreas, dogfish muscle, oyster tissue and bovine liver). The recommended procedure avoids plasma instability, carbon deposit on the cones and does not require sample digestion.
Keywords:ICP-MS   Biological samples   Formic acid solubilization   Trace elements   Electrothermal vaporizer
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号