Comparison of quantification strategies for one-point standard addition calibration: The homoscedastic case
Authors:Richard J.C. Brown  Thomas P.S. Gillam
Affiliation:1. Analytical Science Division, National Physical Laboratory, Teddington, Middlesex, TW11 0LW, UK;2. Department of Physics, Cavendish Laboratory, Cambridge, CB3 0HE, UK
Abstract:Two quantification strategies for one-point standard addition calibration have been compared mathematically. One strategy involved the extrapolation of measurement points to their intercept with the x-axis to determine the analyte content in the unknown sample, and the other strategy is based upon direct calculation of the analyte content in the unknown sample using the instrumental responses obtained during measurement. The cases of both conventional standard addition calibration (C-SAC) and sequential standard addition calibration (S-SAC) have been considered. The homoscedastic situation has been considered, where the absolute precision of the instrumental response is constant. It has been determined that the precision ratio of the two strategies is dependent on surprisingly simple parameters: such as the sample to standard mass ratio (for C-SAC) and the analyte content ratio (for S-SAC).
Keywords:Standard addition   Calibration   Precision   Extrapolation   Comparison