Flow injection calibration methods for atomic absorption spectrometry
Authors:J.F. Tyson  J.M.H. Appleton  A.B. Idris
Affiliation:Department of Chemistry, Loughborough University of Technology, Loughborough, Leicestershire LE 11 3TU Gt. Britain
Abstract:The use of an atomic absorption spectrometer as a detector in flow injection analysis is briefly reviewed. A new simplified model is described for the dispersion effects observed with such systems; the model is based on considering the dispersion to be due to a single hypothetical mixing chamber located immediately prior to the measurement stage. The utility of this approach is demonstrated for two methods of calibration commonly used in atomic absorption spectrometry, and it is shown that flow injection sample and standard handling techniques are comparable to manipulation with volumetric apparatus. The flow injection method has a number of advantages for the analogue of the standard addition method. The use of an exponential concentration gradient is proposed as a novel method of calibration using a single concentrated standard. Results are presented for the determination of chromium in standard steels.
Keywords:
本文献已被 ScienceDirect 等数据库收录!