首页 | 本学科首页   官方微博 | 高级检索  
     


The determination of uranium in aqueous samples by means of a pulsed-source fluorescence spectrometer
Authors:A.T. Rhys Williams  J.N. Miller
Affiliation:Perkin-Elmer Ltd., Beaconsfield, Bucks HP9 1QA Gt. Britain;Department of Chemistry, University of Technology, Loughborough LE11 3TU Gt. Britain
Abstract:Luminescence from aqueous uranyl ions is examined by means of a fluorescence spectrometer with a pulsed xenon light source. Background fluorescence is reduced by using time-based discrimination of the uranyl emission, but interference can still occur from quenchers such as iron(III). Such interferences are reduced by extraction of the uranyl ion into hexane containing trin-n-butyl phosphate, with back-extraction into dilute phosphoric acid before measurement. A detection limit of 5 ng ml?1 is found with a linear calibration range of 0–10 μg ml?1.
Keywords:
本文献已被 ScienceDirect 等数据库收录!