Linear titration plots with ion-selective electrodes |
| |
Authors: | Ivaska A |
| |
Affiliation: | Department of Analytical Chemistry, Abo Akademi, 20500 Abo 50, Finland. |
| |
Abstract: | A computational method for simultaneous determination of the equivalence volume and the Nernstian slope of an ion-selective electrode is described. Treatment of data from linear titration and standard addition procedures is simplified by the use of equal additions of standard solution. Equivalence volumes and concentrations were determined with errors of 0.5-2% by the proposed method. |
| |
Keywords: | |
本文献已被 ScienceDirect PubMed 等数据库收录! |
|