首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Use of photochemical reactions in flow injection: determination of oxalate in urine
Authors:L E Leon  A Ríos  M D Luque de Castro  M Valcárcel
Institution:Department of Analytical Chemistry, Faculty of Sciences, University of Córdoba, Spain.
Abstract:The use of photochemical reactions in flow injection (FI) is reported. The irradiation of an FI reactor with a suitable source facilitates the development of the iron(III)-oxalate reaction, allowing the amperometric determination of the anion in the range 1.0-13.0 micrograms ml-1, with a relative standard deviation of 1.1% and a sampling frequency of 40 h-1. The proposed method was applied successfully to the determination of oxalate in urine samples.