首页 | 本学科首页   官方微博 | 高级检索  
     检索      

基于合成波长法的飞秒激光外差干涉测距方法
引用本文:廖磊,易旺民,杨再华,吴冠豪.基于合成波长法的飞秒激光外差干涉测距方法[J].物理学报,2016,65(14):140601-140601.
作者姓名:廖磊  易旺民  杨再华  吴冠豪
作者单位:1. 清华大学精密仪器系, 精密测试技术及仪器国家重点实验室, 北京 100084; 2. 北京卫星环境工程研究所, 北京 100094
基金项目:国家自然科学基金(批准号:61575105)、清华大学自主科研项目(批准号:20151080460)、航天器高精度测量实验室基金和瞬态光学与光子技术国家重点实验室开放基金(批准号:SKLST201406)资助的课题.
摘    要:本文提出了一种基于合成波长法的飞秒激光外差干涉测距方法.系统采用两个带通滤波器产生两个具有一定波长差的单波长,用于产生合成波长.本方法结构简单,能量利用率高.与双频激光干涉仪在40 mm范围内的比对结果表明,该方法比对残差的标准差为91 nm.

关 键 词:频率梳  外差干涉  绝对距离测量  合成波长法
收稿时间:2016-03-01

Synthetic-wavelength based absolute distance measurement using heterodyne interferometry of a femtosecond laser
Liao Lei,Yi Wang-Min,Yang Zai-Hua,Wu Guan-Hao.Synthetic-wavelength based absolute distance measurement using heterodyne interferometry of a femtosecond laser[J].Acta Physica Sinica,2016,65(14):140601-140601.
Authors:Liao Lei  Yi Wang-Min  Yang Zai-Hua  Wu Guan-Hao
Institution:1. State Key Laboratory of Precision Measurement Technology and Instruments, Department of Precision Instrument, Tsinghua University, Beijing 100084, China; 2. Beijing Institute of Spacecraft Environment Engineering, Beijing 100094, China
Abstract:Large-scale and high precision absolute distance measurement is essential in aerospace technology and advanced manufacturing. Traditional method of measuring distance cannot meet this requirement. Since the advent of optical frequency comb, it has brought a revolutionary breakthrough to absolute distance measurement. In the past decade, there were proposed many methods to measure long absolute distances with high accuracy. Especially, the simple method of using adjacent pulse-to-pulse distance as a ruler for distance measurement has been widely used. The accuracy of this method depends mainly on the knowledge of relative positions of the two overlapped pulses, i.e., pulse-to-pulse alignment. In our previous study, we have proposed a heterodyne interferometer based on synthetic wavelength method with femtosecond laser. The synthetic wavelength is derived from the virtual second harmonic and the real second harmonic, and the real second harmonic is produced by a piece of periodically poled LiNbO3 (PPLN) crystal. However, the second harmonic generation system makes the system complicated, and causes a great optical energy loss. In order to solve this problem, we generate the synthetic wavelength by two spatial band-pass filters in our present study, which can simplify the system greatly. Moreover, we can reduce the optical energy loss and tune the synthetic wavelength by controlling the angle of the filter. The synthetic wavelength used in the present system is 71.39 μm. The interferometric phase of the synthetic wavelength is used as a mark for the pulse-to-pulse alignment. In order to reduce the influences of air disturbance and temperature variation, we set up a thermal-insulated cover for the interferometer to stabilize the environment in the system. By using this cover, the optical path length difference of the system in 450 s can be reduced from 8.56 μm to 0.21 μm. To demonstrate the efficacy of the method described above, the target mirror is moved by eight steps in steps of 5 mm. We compare the measurement results with those obtained by a commercial interferometer, and the residual error is less than 100 nm. Since the measurement range is larger than our previous study, the relative accuracy is better than the previous system. In conclusion, we demonstrate a synthetic-wavelength based absolute distance measurement by using heterodyne interferometry of a femtosecond laser. Two spatial band-pass filters are used to generate the synthetic wavelength, which can simplify the system. The comparison results show that the system has an accuracy better than 100 nm in a displacement of 40 mm. The accuracy of the experimental system can be further improved by making the common-path of the two interferometers longer, locking the fceo to the atomic clock and sampling the data synchronously.
Keywords:optical frequency comb  heterodyne interferometry  absolute distance measurement  synthetic wavelength
本文献已被 CNKI 等数据库收录!
点击此处可从《物理学报》浏览原始摘要信息
点击此处可从《物理学报》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号