Sequencing peptides by electrospray ion-trap mass spectrometry: A useful tool in synthesis of Axinastatin 3 |
| |
Authors: | Chenxi Jia Wei Qi Zhimin He Bin Qiao |
| |
Affiliation: | (1) Chemical Engineering Research Center, School of Chemical Engineering and Technology, Tianjin University, Tianjin, 300072, People’s Republic of China |
| |
Abstract: | Axinastatin 3 as a potential anticancer agent was synthesized by chemical methods. In an electrospray ion-trap mass spectrometer, using one stage of tandem mass spectrometry (MS/MS), the linear peptide intermediate was sequenced via the complementarities of y and b ions. Then, using multistep MS/MS (to MS6), the cyclic peptide was sequenced through sequentially removing one amino acid residue in each stage of MS/MS. The difference of the fragmentation mechanisms and the sequencing approaches between them is discussed. |
| |
Keywords: | Tandem mass spectrometry cyclic peptides fragmentation pathways collisionally activated decomposition |