Abstract:Standard calibration methods used to determine trace analytes usually yield significant deviations from the actual analyte value in the presence of interferents in the assay media. These deviations become of particular concern when the concentration of the analyte is low, and when the results are used to draw mechanistic or kinetic conclusions, for instance in enzyme structure-function studies. In these circumstances, the H-point standard addition method (HPSAM) provides superior precision and accuracy. This method is developed here for the case of the spectrophotometric Griess assay used to determine nitrite in various enzymology investigations, such as nitrite determination in studies of nitrite reductases (NiR), or when determining nitrite as a breakdown product of nitric oxide synthesized by NOS enzymes. The results obtained by HPSAM are contrasted with those of the traditional calibration method.Electronic Supplementary Material Supplementary material is available for this article at
Keywords:Nitrite  Determination  NOS  Enzymes  Griess  Assay