Determination of the acaricide fenbutatin oxide in water samples by automated headspace-SPME-GC/MS |
| |
Authors: | Devos Christophe Moens Luc Sandra Pat |
| |
Affiliation: | Department of Organic Chemistry, Laboratory of Separation Sciences, Ghent University, Krijgslaan 281 S4, B-9000 Ghent, Belgium. |
| |
Abstract: | The analysis of the acaricide fenbutatin oxide (FBTO) having a molecular weight of 1052.66 g mol(-1) in water samples by capillary GC/MS after in-situ derivatization with sodium tetraethylborate (NaBEt4) and headspace-SPME enrichment is described. Automated SPME is performed at 80 degrees C for 30 min. Detection is carried out in the ion monitoring mode with deuterated triphenyltin (TPhTd15) as internal standard. Good linearity (R2 = 0.9993) was obtained in the dynamic range 20 to 1000 ng L(-1) with a limit of detection of 16 ng L(1) (LOD at 3 S/N) and a limit of quantitation of 50 ng L(-1) (LOQ at 10 S/N). Intra-day RSD% for n=6 was 8.9 at the LOQ level. |
| |
Keywords: | Capillary GC/MS Automated headspace‐SPME Derivatization Fenbutatin oxide Water samples |
本文献已被 PubMed 等数据库收录! |