(1) The Connecticut Agricultural Experiment Station, P.O. Box 1106, 06504 New Haven, CT, USA;(2) Present address: Upjohn, Fine Chemicals Division, North Haven, Conn., USA
Abstract:
Summary An empirical method of extrapolating and interpolating gas chromatographic retention times obtained at three equally spaced
isothermal temperatures is described. The accuracy of the method was evaluated from retention time data obtained using packed
glass columns. A procedure for constructing retention time tables for homologous series and the derivation of an equation
for calculating retention times as a function of temperature is also presented.