首页 | 本学科首页   官方微博 | 高级检索  
     


Noise properties of a NMR transceiver coil array
Authors:Pinkerton Robert G  Barberi Enzo A  Menon Ravi S
Affiliation:Department of Medical Biophysics, University of Western Ontario, Ont., Canada.
Abstract:The use of multiple radiofrequency (RF) surface coil elements has applications in both fast parallel imaging and conventional imaging techniques. Through implementation of a simple magnetic decoupling network, 50 Omega matching can be achieved in both the transmitter and receiver chains, enabling the use of conventional RF power amplifiers and preamplifiers for transceive applications. Unlike phased array coil arrangements using low impedance preamplifiers for decoupling, the noise correlation between 50 Omega coils decoupled with discrete components has not been characterized. We have measured the dependence of coil quality factor (Q-factor) and noise correlation on coil separation and shown these quantities to be consistent with theoretical arguments, at least at 4 T (170 MHz). Our results suggest that a coil system for transmission and reception of NMR signals with 50 Omega coils can be built to take advantage of all the benefits of conventional array coils and with the added advantages of using conventional amplifiers.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏