首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Disposable Carbon Electrodes for Liquid Chromatographic Detection of Catecholamines in Blood Plasma Samples
Authors:Pranathi R Perati  Jun Cheng  Petr Jandik  Valoran P Hanko
Institution:Dionex Corporation, 1228 Titan Way, Sunnyvale, California 94088‐3603, USA
Abstract:This paper describes a comparative evaluation of disposable carbon electrodes and conventional glassy carbon electrodes. The detection of catecholamines was demonstrated in processed blood plasma and in the presence of catecholamine metabolites. Calibration plots of norepinephrine, epinephrine, 3,4‐dihydroxybenzylamine and dopamine were linear over three to four orders of magnitude with detection limits of 0.50, 0.73, 1.06, and 1.13 pg, respectively. The relative standard deviation of peak areas was ±2.3% for norepinephrine and ±5.3% for epinephrine from 30 injections of a 10 ng/mL mixed standard. Spike recoveries for norepinephrine, epinephrine, and dopamine from human blood plasma were 86±6%, 81±5%, and 77±4%, respectively.
Keywords:Liquid chromatography  Electrochemical detection  Disposable electrode  Plasma  Catecholamine
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号