Voltammetry of Benzo[a]pyrene in Aqueous and Nonaqueous Media: Adsorptive Stripping Voltammetric Determination at Pencil Graphite Electrode
Authors:Ertuğrul Keskin  Yavuz Yardım  Zühre Şentürk
Affiliation:Yüzüncü Y?l University, Faculty of Science and Letters, Department of Analytical Chemistry, 65080?Van, Turkey
Abstract:A detailed study of the electrochemical oxidation of Benzo[a]pyrene (BaP) at the glassy carbon and pencil graphite electrodes was carried out in aqueous and nonaqueous media. Using square‐wave stripping mode, the compound yielded a well‐defined voltammetric response at pencil graphite electrode in acetate buffer, pH 4.8 at +1.13 V (vs. Ag/AgCl) (a preconcentration step being carried out at a fixed potential of +0.70 V for 180 s). The process could be used to determine BaP concentrations in the range 0.25–1.25 μM, with a detection limit of 0.027 μM (6.82 μg L?1). The applicability to assay of spiked human urine samples was also illustrated.
Keywords:Benzo[a]pyrene (BaP)  Cyclic voltammetry  Adsorptive stripping square‐wave voltammetry  Nonaqueous and aqueous media  Urine  Glassy carbon electrodes  Pencil graphite electrodes