Highly Sensitive Electrochemical Sensor for Determination of Vitamin D in Mixtures of Water‐Ethanol
Authors:Fernando H Cincotto  Thiago C Canevari  Sérgio A S Machado
Institution:Institute of Chemistry, State University of S?o Paulo, PO Box 780, 13560‐970 S?o Carlos, SP, Brazil tel/fax: (+55)?16?3521–3109
Abstract:This paper describes the electrochemical determination of vitamin D2 (ergocalciferol) and D3 (cholecalciferol) in mixed organic/water solvent, using a glassy carbon electrode (GCE). The mixing ratio of organic/water solvent has an important influence on the electrocatalytic response of D vitamins on the surface of the glassy carbon electrode. Well‐defined peaks for Vitamin D2 and D3 were observed in a 40 % ethanol/60 % water solution with lithium perchlorate as the support electrolyte. This study demonstrated that the glassy carbon electrode is highly sensitive for the determination of vitamin D2 and D3, with a limit of detection of 0.13 and 0.118 µmol L?1, respectively. No significant interference was seen for vitamins A, E and K in the detection of vitamin D.
Keywords:Vitamin D  Electrochemical sensors  Organic/water solution  Electroanalysis