首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of silver in fresh water by atomic absorption spectrometry following flotation preconcentration by iron(III) collectors
Authors:Katarina Čundeva  Trajče Stafilov
Affiliation:Institute of Chemistry, Faculty of Science, St. Cyril and Methodius University, Skopje, Macedonia, XX
Abstract:Colloid precipitate flotation of silver from fresh water is applied for preconcentration and separation. Optimal conditions using hydrated iron(III) oxide and iron(III) tetramethylenedithiocarbamate as collectors were investigated. Various factors affecting the silver recovery, including collector mass, nature of the supporting electrolyte, pH of the working medium, electrokinetic potential of the collector particle surfaces, type of surfactant, induction time etc., were checked. Within the optimal pH range (5.5–6.5) silver was separated quantitatively (94.9– 100.0%) with 30 mg Fe(III) as collector. The content of silver was determined by electrothermal atomic absorption spectrometry and compared to that from inductively coupled plasma-atomic emission spectrometry. The detection limit of silver by the method described is 0.01 μg/L.
Keywords:
本文献已被 SpringerLink 等数据库收录!