首页 | 本学科首页   官方微博 | 高级检索  
     


Coupling micelle-mediated extraction using mixtures of surfactants and fluorescence measurements with a fiber-optic for the screening of PAHs in seawater
Authors:Delgado Barbara  Pino Veronica  Ayala Juan H  Gonzalez Venerando  Afonso Ana M
Affiliation:Department of Analytical Chemistry, Nutrition and Food Science, University of La Laguna, Campus de Anchieta, Astrofisico Francisco Sanchez s/n, E-38205 La Laguna, Spain.
Abstract:The screening of PAHs from seawater samples using cloud-point extraction (CPE) as a step prior to their determination by fluorescence measurements with a fiber-optic is proposed. The CPE is carried out with the nonionic surfactants mixture POLE and Brij 30. The fluorescence measurement parameters were optimized, allowing selection of benzo(a)pyrene (B(a)Py) and benzo(k)fluoranthene (B(k)Ft) as the target analytes for the screening. The reproducibility of the whole screening system, expressed as relative standard deviation, was 9.0% for B(a)Py and 12.1% for B(k)Ft (both for n = 7). The reliability of the method was established at five concentrations for B(a)Py (between 0.5 and 3.3 times the detection limit: 0.31 ng ml(-1)) and at three concentrations for B(k)Ft (between 0.6 and 2.5 times the detection limit: 0.56 ng ml(-1)). The resolution of binary mixtures of these PAHs at different levels of concentration, and a study of the interferences with the rest of the PAHs were also carried out.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号