首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Determination of cadmium in river water samples by flame AAS after on-line preconcentration in mini-column packed with 2-aminothiazole-modified silica gel.
Authors:Ilton L Alcantara  Paulo S Roldan  Gustavo R Castro  Fabrício V Moraes  Fábio A Silva  Cilene C F Padilha  Jorge D Oliveira  Pedro M Padilha
Institution:Instituto de Química, Departamento de Química Analítica-UNESP, Araraquara, SP, Brasil.
Abstract:A rapid and sensitive method was developed to determine trace levels of Cd2+ ions in an aqueous medium by flame atomic absorption spectrometry, using on-line preconcentration in a mini-column packed with 100 mg of 2-aminothiazol modified silica gel (SiAT). The Cd2+ ions were sorbed at pH 5.0. The preconcentrated Cd2+ ions were directly eluted from the column to the spectrometer's nebulizer-burner system using 100 microL of 2 mol L(-1) hydrochloric acid. A retention efficiency of over 95% was achieved. The enrichment factor (calculated as the ratio of slopes of the calibration graphs) obtained with preconcentrations in a mini-column packed with SiAT (A = -1.3 x 10(-3) + 1.8 x 10(-3)Cd2+]) and without preconcentrations (A = 4 x 10(-5) + 3.5 x 10(-5)Cd2+]), was 51 and the detection limit calculated was 0.38 microg L(-1). The preconcentration procedure was applied to determine trace levels of Cd in river water samples. The optimum preconcentration conditions are discussed herein.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏