Abstract: | An annular helium inductively coupled plasma (He ICP) was generated at atmospheric pressure. No external cooling was used to stabilize the plasma. Aqueous solution was injected into the plasma without any difficulty. Preliminary results revealed that the annular He ICP was capable of exciting elements such as Cl and Br which possess high excitation energies. Atomic emission detection limits for Cl and Br were improved by factors of 63 and 34, respectively, as compared to the results obtained from the argon inductively coupled plasma. The excitation temperature of the annular He ICP (4180 K) was less than that of an Ar ICP (5570 K). |