Supported liquid membrane work-up in combination with liquid chromatography and electrochemical detection for the determination of chlorinated phenols in natural water samples |
| |
Authors: | M. Knutsson L. Mathiasson J. Å. Jönsson |
| |
Affiliation: | (1) Department of Analytical Chemistry, University of Lund, P.O. Box 124, 221 00 Lund, Sweden |
| |
Abstract: | Summary A sample preparation method has been developed for the determination of chlorinated phenols in water. The method is based on a supported liquid membrane extraction system connected on-line to liquid chromatography with electrochemical detection. The supported liquid membrane technique utilizes a porous PTFE membrane. The membrane is impregnated with an organic solvent which forms a barrier between two aqueous phases and enables selective extraction. The technique can easily be coupled in a flow system. In this investigation five chlorinated phenols (1–5 chlorine atoms) were extracted from natural water samples. Extraction for 30 minutes resulted in detection limits of approximately 25 ng L–1. |
| |
Keywords: | Column liquid chromatography Liquid membrane extraction Chlorinated phenols Natural water samples |
本文献已被 SpringerLink 等数据库收录! |