Fluorimetric determination of aluminium in serum
Authors:Y Suzuki  S Imai  T Kamiki
Abstract:A convenient fluorimetric method for the routine determination of aluminium in serum has been developed using lumogallion 4-chloro-3-(2,4-dihydroxyphenylazo)-2-hydroxybenzene-1-sulphonic acid]. Losses of aluminium during deproteinisation of the serum were prevented by treatment with a combination of 20 or 30% m/V trichloroacetic acid (TCA) and 5% m/VTCA. Iron(III) was removed by extraction into chloroform with capriquat (methyltrioctylammonium chloride) as an Fe3+-lumogallion-capriquat ternary complex. The interference from Cu2+ was eliminated by using thiosulphate as a masking agent. The detection limit was 3.6 ng ml-1 and the calibration graph was linear up to 1.4 micrograms ml-1 of aluminium. Using the proposed method, the average concentration of aluminium in the serum of healthy subjects was found to be 6.8 ng ml-1, in agreement with values reported in the literature.