Computer controlled diode laser spectrometer with a helium evaporation cryostat and spectroscopy of thev 3 vibration of NCO
Authors:R Brüggemann  M Petri  H Fischer  D Mauer  D Reinert  W Urban
Institution:(1) Institut für Angewandte Physik, Universität, Wegelerstrasse 8, D-5300 Bonn 1, Fed. Rep. Germany;(2) Present address: Fraunhofer-Institut für Physikalische Meßtechnik, D-7800 Freiburg, Fed. Rep. Germany;(3) Present address: I. Physikalisches Institut, Universität, D-5000 Köln, Fed. Rep. Germany
Abstract:A computer-controlled diode laser spectrometer for the 1200 to 2500 cm–1 spectral region is described. The spectrometer has been applied to high resolution spectroscopy of the NCO radical at 5.2 mgrm. The lead-salt diode lasers are cooled to their operating temperature with a temperature adjustable helium evaporation cryostat. Computer-controlled tuning procedures for the frequency tuning of the diode lasers have been developed; they are independent of tables describing the tuning characteristics of the diode lasers. 41 lines of the antisymmetric stretching-vibrationv 3 of the linear NCO radical have been observed. We were able to detect vibration-rotation transitions in both2 pgr 1/2 and2 pgr 3/2 fine structure sublevels. These measurements led to the precise determination of additional molecular constants.
Keywords:42  60  33  20