引用本文:刘葵,耿玉珍.比值光谱导数法同时测定铝合金中铁,铜,锌[J].分析化学,1998,26(10):1201-1204.
作者姓名:刘葵  耿玉珍
作者单位:山东农业大学基础部!泰安271018
摘    要:用比值光谱-导数分光光度法,在pH5.5缓冲溶液中,利用金属-2-(5-溴-2-吡啶偶氮)-5-二乙氨基酚(5-Br-PADAP)-OP三元络合物显色体系,对混合物中铁,铜,锌三组分进行了同时测定。合成试样5次测定回收率在97.3%-104.4%之间。应用于铝合金中铁,铜,锌的测定,各6次测定的RSD分别为3.66%,1.38%,2.03%。

关 键 词:比值导数光谱        铝合金  分析

Simultaneous Determination of Iron. Copper and Zinc in Aluminium Alloy by the Ratio Spectrum-Derivative Spectrophotometry
Liu Kui,Geng Yuzhen,Yang Jingzhi,Sun Yanhua.Simultaneous Determination of Iron. Copper and Zinc in Aluminium Alloy by the Ratio Spectrum-Derivative Spectrophotometry[J].Chinese Journal of Analytical Chemistry,1998,26(10):1201-1204.
Authors:Liu Kui  Geng Yuzhen  Yang Jingzhi  Sun Yanhua
Abstract:Ratio spectro-derivative spectrophotometry was used for the simultaneous determination of Fe, Cu and Zn in aluminium alloy. In pH 5. 5 buffer solution, the overlapped spactra of Me-2-(5-bromo- 2pyridylazo ) -5 - diethylaminophenol (5 - Br- PADAP ) - OP complexes can be resolved by the method for the de termination. The rmovery of synthetic sample is in the range of 97 .3% -104.4% (n = 5). The relative standard deviations for the determination of Fe, Cu, Zn in aluminium alloy were 3. 7%, 1. 4% and 2.0% (n = 6) respectively.
Keywords:Ratio specteum-derivative spectrophotometric method  simultaneous determination  iron  copper  zinc  