Comparison of phenyl-type columns in the development of a fast liquid chromatographic system for eighteen opiates commonly found in forensic toxicology |
| |
Authors: | Dams R Lambert W E Clauwaert K M De Leenheer A P |
| |
Affiliation: | Laboratorium voor Toxicologie, Universiteit Gent, Ghent, Belgium. |
| |
Abstract: | We report a precise and reliable method for the detection of 18 of the most commonly found opiates on the Belgian legal and illicit market, by ion-exchange, reversed-phase high-performance liquid chromatography, using a conventional phenyl-type analytical column (150x4.6 mm I.D., particle size 5 microm) and diode-array detection. We also describe a performance (efficiency and sensitivity) comparison of this column to a recently developed "high-speed" column (53x7.0 mm I.D., particle size 3 microm) packed with the same stationary phase, and used under slightly adjusted flow and gradient conditions. The final method, using the "high-speed" column, showed a significant reduction (55%) in analysis time without loss of resolution and sensitivity. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|