首页 | 本学科首页   官方微博 | 高级检索  
     检索      

一种基于双工作电极-双通道的毛细管电泳电化学检测系统
引用本文:杨冰仪,莫金垣,赖容.一种基于双工作电极-双通道的毛细管电泳电化学检测系统[J].化学学报,2003,61(9):1461-1465.
作者姓名:杨冰仪  莫金垣  赖容
作者单位:1. 中山大学化学与化学工程学院,广州,510275;广东药学院预防医学系,广州,510224
2. 中山大学化学与化学工程学院,广州,510275
基金项目:国家自然科学基金 (No.2 96750 33),广东省自然科学基金 (No.0 0 1 2 37)资助项目
摘    要:报道了一种双工作电极-双通道毛细管电泳电化学检测系统,实现电导和安培 同时检测或者安培与安培检测联用,使两种方法相互补充,发挥各自的优势。其中 ,工作电极与检测池的制作工艺简单,操作简便,通过不锈钢针管和毛细管作为套 管,无需三维微调装置即可简单实现双工作电极的准确放置及分离毛细管与工作电 极的准确对接,并根据分析体系的需要采用不同类型的工作电极和检测器;同时采 用复式滤波电路解决了不同检测器之间的电场叠加对输出信号的干扰问题。采用该 装置可以同时检测复杂体系中的电活性和惰性物质,或同时测定只能氧化或只能还 原的物质,还可以对具有氧化还原性质的物质进行纯度的确证。将该装置应用于实 际样品的测定,节约了分析时间,提高了分析速度,扩大了检测范围,结果令人满 意。

关 键 词:毛细管电泳  电极  电化学检测器
修稿时间:2003年1月26日

A New Dual-electrode and Dual-channel Electrochemical Detection System for Capillary Electrophoresis
Institution:School of Chemistry and Chemical Engineering, Sun Yat-sen University
Abstract:A new dual-electrode and dual-channel electrochemical detection technology for capillary electrophoresis is described, in which two detectors the amperometric detector and the conductometric detector or two conductometric detectors are connected to the same capillary electrophoresis system. The system possesses the advantages of two electrochemical detectors including sparing time, improving the analytical speed and expanding the sample range. The fabrication of the detection cell and the working electrode are convenient. It is easy to assemble and disassemble the working electrode and the separation capillary. With the stainless steel syringe needles and capillary as the guide tubes, the alignment of the working electrodes to the outlet of the separation capillary can be easily done without the aid of three-dimensional adjustor. Different kinds of working electrode and detector can be chosen according to the different analytical aims. By using a laboratory-built filter circuit the signals disturbance generated from different detectors can eliminate effectively. With this system the electroactive (oxidative or reductive) and inert components in a complicated sample can be detected simultaneously. It can also be used to verify analytes. The system is applied to detect real sample with satisfactory results.
Keywords:capillary electrophoresis  dual-channel electrochemical detector  dual-electrode  
本文献已被 CNKI 维普 万方数据 等数据库收录!
点击此处可从《化学学报》浏览原始摘要信息
点击此处可从《化学学报》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号