A photometric flow measurements method for characterisation of HPLC pumps
Authors:K Shoikhet  H Engelhardt
Institution:(1) Angewandte Physikalische Chemie, Universität des Saarlandes, 66123 Saarbrücken, Germany
Abstract:Summary The measurement of flow constancy and pulsation amplitudes of HPLC pump based on the photoconversion of malachite green leucocyanide is described. The irradiation time and hence the degree of conversion of the leucocyanide is correlated to flow fluctuations of piston driven HPLC pumps. It is possible to measure the amplitude of pulsations and determine the constancy of the flow rate. Most of the pumps show a flow stability better than ±1% measured as relative standard deviation of the flow rate under HPLC conditions (pressure drop 100 bar at 1 ml/min flow rate). The most expensive pump of those tested showed less than ±0.5% flow instability, however, this result was achieved by installation of a large-volume pulsation damper. The method described also allows flow rate measurements under FIA conditions where there is little or no pressure drop at the pump outlet.
Keywords:Column liquid chromatography  Pumps  Flow control