首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Electrocardiographic gating and monitoring in NMR imaging
Authors:R E Wendt  R Rokey  G W Vick  D L Johnston
Institution:Department of Radiology, Baylor College of Medicine, Houston, Texas 77030.
Abstract:ECG gating and monitoring during NMR imaging may be achieved reliably by applying the principles in this tutorial. In order to use the ECG signal both for triggering and for patient monitoring it must have a prominent R-wave, while at the same time must have little artifact from gradient switches or the Lorentz voltage across the aorta, and not be significantly distorted by gradient switching artifacts. The twin goals of no image artifacts and minimal ECG artifacts may be achieved by the following means: (1) using ECG electrodes with minimal metal, (2) selecting electrodes and cables with no ferrous metals, (3) placing the limb electrodes close together, (4) placing the line between the limb electrodes and the leg electrode parallel to the magnetic flux lines and, if possible, parallel to the transverse component of the gradient flux lines, (5) keeping the area between the limb electrodes and the leg electrode small, (6) placing that area in the center of the imager and (7) twisting or braiding the cables. Following these principles allows artifact-free images and reliable ECG monitoring during ECG-gated NMR imaging examinations.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号