Adsorption micro calorimeter |
| |
Authors: | V García-Cuello J C Moreno-Pirajan L Giraldo-Gutiérrez K Sapag G Zgrablich |
| |
Institution: | 1. Grupo de Investigación en Sólidos Porosos y Calorimetría, Departamento de Química, Facultad de Ciencias, Universidad de Los Andes, Bogotá, Colombia 2. Departamento de Química, Facultad de Ciencias, Universidad Nacional de Colombia, Bogotá, Colombia 3. Instituto de Física aplicada (INFAP), CONICET, Universidad Nacional de San Luis, San Luis, Argentina
|
| |
Abstract: | In this work, it is described an innovative heat flux micro calorimeter Tian-Calvet type designed to measure adsorption heats and reactions as well as adsorption isotherms. It consists in an adsorption instrument for volumetric gases, which is coupled to the micro calorimeter. The changes in the pressure are monitored by means of high sensitivity and high precision pressure transducers. The micro calorimeter has thermo elements that work by a Seebeck effect, in a twin cells system. The cells are inside a box in which the temperature can be adjusted from 77 to 300 K. The sensitiveness of the calorimeter is established by applying a perfectly known electric work. The results corresponding to the electric calibration, the base line stability determination and the time constant in the equipment are shown. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|