首页 | 本学科首页   官方微博 | 高级检索  
     


Molar mass characterization of cationic methyl methacrylate-ethyl acrylate copolymers using size-exclusion chromatography with online multi-angle light scattering and refractometric detection
Authors:Wittgren Bengt  Welinder Anette  Porsch Bedrich
Affiliation:National Chromatographic R&A Center, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian 116011, China.
Abstract:Size-exclusion chromatography (SEC) combined with online multi-angle light scattering (MALS) and refractometric (RI) detection has been employed for the molar mass characterisation of water-insoluble cationic methyl methacrylate-ethyl acrylate copolymers (Eudragit RS and RL). Due to their positive charge, cationic polymers are particularly difficult to separate on a SEC column, in worst cases being completely adsorbed on the oppositely charged packing material. This work has examined how a careful addition of salt (LiCl) to the copolymer solution in ethanol decreases the electrostatic interactions, clearly seen as a decrease in elution volume from the SEC column as well as an improved recovery. At a certain level of ionic strength, typically about 50 mM, the copolymer recovery from the SEC column reached 100% and molar mass distributions corresponding to the complete sample could be obtained. The combined MALS/RI detection gives the opportunity to measure the absolute molar mass independent of recovery and retention. Thus, in this study, it turned out to be a favourable tool for tracing the changes in elution behaviour of the charged copolymer as the ionic strength was increased.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏