首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of blood longitudinal relaxation time (T1) at high magnetic field strengths
Authors:Dobre Mircea C  Uğurbil Kâmil  Marjanska Malgorzata
Affiliation:Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, MN 55455, USA.
Abstract:In this study, a circulation system was used to measure T(1) values of bovine blood under physiological conditions at field strengths of 4.7, 7 and 9.4 T. Results show that T(1) increases linearly with magnetic field B(0) and can be described with the equation T(1)=129 ms/T B(0)+1167 ms for magnetic field strengths between 1.5 and 9.4 T.
Keywords:Longitudinal relaxation   T1   Blood   High magnetic field
本文献已被 ScienceDirect PubMed 等数据库收录!