Effect of the encapsulating material on the peak3/peak5 response ratio of TLD-300 irradiated with neutrons of various energy
Authors:M Angelone  P Batistoni  M Pillon
Institution:

Associazione EURATOM-ENEA sulla Fusione Centro Ricerche Frascati, C.P. 65, I-00044 Frascati, Rome, Italy

Abstract:The analysis of experimental data available in the literature shows that the neutron sensitivity ratio m=k3/k5 for the two main glow peaks P3 and P5 of TLD-300 (for ribbons of 3.2×3.2×0.9 mm3) is dependent upon the response ratio R3/R5 of these two peaks. This finding is in agreement with predictions from an equation derived from the two-peak method equations. The result implies that the peak response ratio R3/R5 for TLD-300 surrounded by Tissue Equivalent material is independent of the encapsulating material. This is also demonstrated experimentally.
Keywords:Mixed field dosimetry  Thermoluminescence  Neutron sensitivity