Coupling protocol for thin layer chromatography/matrix-assisted laser desorption ionization |
| |
Authors: | J T Mehl A I Gusev D M Hercules |
| |
Institution: | (1) Department of Chemistry, Vanderbilt University, 37235 Nashville, TN, USA |
| |
Abstract: | Summary This paper reports the investigation and optimization of the thin layer chromatography/matrix-assisted laser desorption/ionization
(TLC-MALDI) coupling protocol. The fundamental coupling parameters which influence sensitivity and lateral analyte spreading
are extraction solvent selection, extraction time, and pressure. Selection of the solvent is dependent upon its extraction
efficiency, which has been correlated with extraction solventR
f value. Maximum extraction efficiency occurs after two minutes for the analyte/solvent system studied. High extraction efficiency
solvents cause significant lateral spreading of analyte; up to a three-fold increase in initial analyte spot size. Analyte
recovery was found to be limited by the silica gel inter-partice porosity and the solvent extraction efficiency. For maintaining
chromatographic resolution and maximizing signal intensity, extraction solvents withR
f values between 0.4 to 0.6 are optimal. The upper analyte recovery limit, using extraction solvents within thisR
f range, is estimated at 22%.
Dedicated to Professor Werner Haerdi on the occasion of his 70th birthday. |
| |
Keywords: | Thin layer chromatography (TLC) MALDI TLC-MS |
本文献已被 SpringerLink 等数据库收录! |
|