首页 | 本学科首页   官方微博 | 高级检索  
     


Self-calibration of a phase-shifting adapter for Fizeau interferometers
Authors:Yuan Ouyang  Mang Ou-Yang  Hou-Cheng Chou
Affiliation:(1) Department of Electrical Engineering, Chang Gung University, 259 Wen-Hwa 1st Road, Kwei-Shan, Taoyuan County, 33302, Taiwan;(2) Department of Optics and Photonics, National Central University, No. 300, Jungda Rd, Jhongli City, Taoyuan County, 32001, Taiwan
Abstract:Phase-shifting interferometers with three piezoelectric transducers (PZTs) on their phase-shifting adapter (PSA) are widely used in precision measurement. The guarantee of the PSA to stretch as a flat is very important to the accuracy of the measurement. In this paper, we propose a systematic self-calibration method to calibrate the different displacements of the PZTs so that the PSA can be moved evenly as a flat. The proposed method uses only the interference patterns to calibrate the three PZTs and information about the exact position of the PZTs and the instrument parameters of the interferometer are not required. Experimental results show that, after calibration by the proposed method, the deviation of the measured phase shift error could be reduced about 43.1 times from 3.492 to 0.081°.
Keywords:calibration  interferometry  phase-shifting adapter (PSA)  piezoelectric transducer
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏