首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Automated at-column injection into narrow bore capillary GC columns
Authors:Gunnar Hagman  Johan Roeraade
Abstract:An injector designed for automatic direct liquid injection into narrow bore capillary GC columns has been constructed and evaluated. The tip of the syringe needle is aligned with, and positioned close to, the column entrance in a small, pressurized cavity: when the sample is dispensed it is immediately forced into the column by the action of the surrounding carrier gas. A standard autosampler equipped with a standard stainless steel syringe needle was utilized for at-column sample transfer into 100 μm i.d. columns. RSD values for n-alkanes were between 0.1 and 0.3% for relative area counts and approximately 1% for absolute area counts.
Keywords:Gas chromatography (GC)  Narrow bore capillary columns  Sample introduction  Automated at-column injection