首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Chromium determination in sea water by electrothermal atomic absorption spectrometry using Zeeman effect background correction and a multi-injection technique
Authors:P Bermejo-Barrera  Jorge Moreda-Pi?eiro  Antonio Moreda-Pi?eiro and Adela Bermejo-Barrera
Institution:(1) Department of Analytical Chemistry, Nutrition and Bromatology, Faculty of Chemistry, University of Santiago de Compostela, Avenida de las Ciencias, s/n. E-15706-Santiago de Compostela, Spain, ES
Abstract:Electrothermal atomic absorption spectrometry (ETAAS) applying a Zeeman effect background correction system (ZEBC) and a tranverse heated atomizer was used to directly determine chromium in sea water. Calcium chloride (at a concentration of 20 mg L–1) was applied as chemical modifier with optimum charring and atomization temperatures of 1600°C and 2000°C, respectively. The detection limit was 0.2 μg L–1, by injecting 20 μL aliquot of sea water sample. This detection limit could be reduced further to 0.05 μg L–1, using multiple injections (injection of five 20 μL aliquot of sea water). The accuracy of the methods developed were confirmed by analyses of different certified reference materials. Finally, interferences from major and minor components of sea water are studied. Received: 20 February 1997 / Revised: 26 May 1997 / Accepted: 8 June 1997
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏