首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Direct complexometric determination of magnesium in the presence of uranium, iron and aluminium
Authors:Mareska V
Institution:Nuclear Research Institute CAV Rez near Prague, Czechoslovakia.
Abstract:Conditions for the direct complexometric determination of magnesium in the presence of uranium, iron and aluminium are described. The method is based on the masking of uranium by hydroxylamine, and of iron and aluminium by triethanolamine. Methylthymol Blue is used as indicator for the EDTA titration of magnesium at pH 10.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏