Stripping voltammetric determination of trace levels of uranium by synergic adsorption |
| |
Authors: | Marina Mlakar Marko Branica |
| |
Institution: | Centre for Marine Research Zagreb, Rudjer Bo?kovi? Institute, Box 1016, 41001-Zagreb, Croatia Yugoslavia |
| |
Abstract: | A voltammetric method is reported for the determination of low uranium concentrations in natural waters. A very strongly adsorbed mixed complex of uranyl ion with 2-thenoyltrifluoroacetone and tri-n-butyl phosphate is applied. Synergic adsorption is obtained compared with the adsorption obtained with either ligand alone. In 0.55 mol dm?3 sodium chloride, the detection limit was 10?10 mol dm?3 uranyl ion after accumulation for 10 min at ?0.1 V vs. Ag/AgCl. For the natural level of uranium in sea water (about 10?8 mol dm?3), accumulation for 2 min was sufficient. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|