Analytical possibilities of square-wave amperometry |
| |
Authors: | Adam Hulanicki Wojciech Jedral |
| |
Affiliation: | Department of Chemistry, University of Warsaw, Warsaw Poland |
| |
Abstract: | The application of improved electronic equipment for square-wave amperometry (SWA) is described. This equipment has better resolution between waiting and sampling times and a simple, low-noise logic part. Titrations with electrogenerated bromine, iodine, iron(II) and silver ions were investigated as examples to compare SWA with d.c. amperometry (DCA). The sensitivity of SWA in titrations of < 10?5 moll?1 concentrations with iodine or silver(I) is more than 100 times greater than that of DCA. For measurements of higher concentations, DCA is preferred because of its better linear response. The change in current near the end-point of titrations with silver ions is sharper with SWA than with DCA. |
| |