Flow-injection and liquid chromatographic determination of aluminum based on its fluorimetric reaction with 8-hydroxyquinoline-5-sulphonic acid in a micellar medium |
| |
Authors: | José Ignacio Garciá Alonso Angeles López Garciá Alfredo Sanzmedel Elisa Blanco Gonzales Les Ebdon Phil Jones |
| |
Affiliation: | Departmento de Química Física y Analítica, Facultad de Química, Universidad de Oviedo, Julian Clavería s/n, Oviedo Spain;Department of Environmental Sciences, Polytechnic Southwest, Drake Circus, Plymouth PL4 8AA Great Britain |
| |
Abstract: | Aluminum reacts with 8-hydroxyquinoline-5-sulphonic acids (HQS) in cationic micelles of cetyltrimethylammonium bromide (CTAB) to form a strongly fluorescent compound. In the micellar media the formation of the Al—HQS complex is very fast, allowing a useful continuousdetermination of aluminum in flow systems. At pH 6.3, for 2 × 10-4 M HQS and 2 × 10-3 M CTAB the detection limit for a luminum by flow injection is 0.1 ng ml-1. The calibration is linear up to 100 ng ml-1 and the peak-height precision is 2% at the 10 ng ml-1 level. Interfences are greatly decreased compared with the batch method. The reaction has also been applied to the post-column detection by ion chromatography of aluminium in fresh and high-salinity wates and for speciation of aluminium in blodd serum after separation of serum proteins by ion-exchange liquid chromatography. |
| |
Keywords: | |
本文献已被 ScienceDirect 等数据库收录! |
|