Enzymatic determination of urea in undiluted whole blood by flow injection analysis using an ammonium ion-selective electrode
Authors:Bo A Petersson
Institution:Radiometer A/S, Emdrupvej 72, DK-2400 Copenhagen Denmark
Abstract:A flow-injection system is described for the assay of urea in undiluted whole blood. Urea is quantified by means of an ammonium ion-selective electrode covered with a membrane with covalently immobilized urease. The enzymatically generated ammonium ion is directly related to the urea concentration. Interference from potassium is reduced by adjusting the potassium ion concentration in the carrier stream and in the aqueous calibration solutions to 4.0 mM; it can be eliminated by measuring the potassium ion concentration in the sample separately and applying a mathematical correction for the K+ contribution to the signal. The linear measuring range is 1–40 mM urea, with an injection frequency of 40 h?1 and a standard deviation of 1% for whole blood samples. The result of the measurement is obtained within 25 s from the time of injection. Vatiations in the hematocrit level of the sample have no effect on the measurement. The results obtained by the flow-injection method are in excellent agreement with those found routinely at a local hospital. The sensor is stable for more than 25 days.