Effects of ionic radiological and chemical interferences on the chromatographic separation of a radionuclide standard solution |
| |
Authors: | J E Roane T A DeVol J D Leyba R A Fjeld |
| |
Institution: | (1) Department of Environmental Engineering and Science, Clemson University, 29634 Clemson, SC, USA;(2) Savannah River Technology Center, Westinghouse Savannah River Co., 29808 SC, Aiken, USA |
| |
Abstract: | For characterizing radioactive samples there are two major considerations in the application of a coupled liquid chromatography
and on-line scintillation counting system: (1) radiological interferences and (2) chemical interferences from the matrix.
A study was conducted to identify which interferences from typical matrices found at several Department of Energy facilities
affected the separation of a radioactive tracer solution by the coupled system. The selection of potential interferent materials
was determined through a review of characterization and monitoring studies of surface water, ground water, and high level
waste tank supermatant at those facilities. Incremental mass loadings of contaminant were mixed with a standardized radioactive
tracer (55Fe,63Ni,90Sr and147Pm) and then injected into a coupled system. The resultant chromatograms were compared to the chromatogram of the standard
radionuclide solution to determine the effects of the chemical or radiological constituent. Relative to the radionuclide solution,137Cs was the only activation/fission product used in this study to effect a radiological interference. For the natural uranium
series, a radiological interference was observed for90Sr due to either a uranium isotope or a decay product of the series. No rad interference was observed from228Th, though it must be noted that the elution program was not capable of completely separating the decay series nuclides of
natural uranium or thorium. For the chemical interferences, the effects are twofold since the chemical can affect the concentration
of ions on the pre-concentration stage as well as the chromatographic separation. The general trend observed was that increasing
the ionic strength of the chemical resulted in decreased retention fractions on the pre-concentration column and significant
shifts in peak elution times. |
| |