首页 | 本学科首页   官方微博 | 高级检索  
     


The correct and incorrect way to calibrate a Compton suppression counting system for gamma-ray efficiency
Authors:S. Landsberger  S. Biegalski  R. Kapsimalis
Affiliation:1. Nuclear Engineering Teaching Laboratory, University of Texas at Austin, R-9000, Austin, TX, 78712, USA
Abstract:Gamma-ray efficiency calculations for a germanium detector have been made for a Compton suppression system. Results have shown that for radionuclides that have gamma rays in coincidence the photopeaks can be severely depressed leading to erroneous results. While this can be overcome in routine neutron activation analysis using a comparator method, special consideration must be given to determine the suppression for coincident gamma rays when calculating the efficiency curve and for radionuclide activities. This is especially important for users of the k 0 method and for fission product identification using Compton suppression methods.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号