首页 | 本学科首页   官方微博 | 高级检索  
     


Catalytic Adsorptive Stripping Voltammetric Procedure for Determination of Total Chromium in Environmental Materials
Abstract:A sensitive catalytic adsorptive stripping voltammetric procedure for determination of traces of total chromium in environmental samples is reported. The method is based on the preconcentration of a Cr(III)? H2DTPA complex by adsorption at the HMDE from an acetate buffer solution at the potential ?1.0 V vs. Ag/AgCl. Total chromium was determined as Cr(III) after reduction of Cr(VI) to Cr(III) by NaHSO3. In order to stabilize the signal of Cr(III) the measurements were performed at 5 °C. The calibration graph for chromium for an accumulation time of 60 s was linear in the range from 5×10?10 to 5×10?8 mol L?1. The relative standard deviation for a chromium concentration of 1×10?8 mol L?1 was 3.9% (n=5). The detection limit for accumulation time of 60 s was about 8×10?11 mol L?1. The validation of the procedure was performed by the analysis of the certified reference materials.
Keywords:Chromium  Determination  Stripping analysis  DTPA
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号