Electro‐oxidation and Determination of Tripelennamine Hydrochloride at MWCNT‐CTAB Modified Glassy Carbon Electrode |
| |
Authors: | Jayant I. Gowda Sharanappa T. Nandibewoor |
| |
Affiliation: | 1. P.?G. Department of Studies in Chemistry, Karnatak University, Dharwad‐580 003, India, Tel.: +91?836?2770524;2. Fax: +91?836?2747884 |
| |
Abstract: | An electrochemical method for the determination of tripelennamine hydrochloride (TPA) using cetyltrimethylammoniumbromide‐multiwalled carbon nanotubes modified glassy carbon electrode (MWCNT‐CTAB/GCE) was developed. Because of good electrical conductivity of MWCNT and catalytic behavior of CTAB, new electrode significantly enhances the sensitivity for the detection of TPA. Parameters such as amount of modifier suspension, scan rate, pH of measure solution, heterogeneous rate constant were investigated. The electrode exhibits a linear potential response in the range of 1.0×10?8 M to 3.0×10?6 M with a detection limit of 2.38× 10?9 M. The modified electrode was successfully applied to the determination of TPA in pharmaceutical and real samples. |
| |
Keywords: | Tripelennamine hydrochloride Electro‐oxidation Pharmaceutical analysis Detection limit |
|
|