Development and validation of a sensitive LC-MS/MS method with electrospray ionization for quantitation of pramipexole in human plasma: application to a clinical pharmacokinetic study |
| |
Authors: | Bharathi D Vijaya Hotha Kishore Kumar Sagar P V Vidya Kumar S Sirish Naidu A Mullangi Ramesh |
| |
Institution: | Bioanalytical Department, Integrated Product Development, Dr Reddy's Laboratories Ltd, Bachupalli, Hyderabad-500 072, India. |
| |
Abstract: | A highly sensitive and specific LC-MS/MS method has been developed and validated for the estimation of pramipexole (PPX) with 500 microL human plasma using memantine as an internal standard (IS). The API-4000 was operated under multiple-reaction monitoring mode (MRM) using the electrospray ionization technique. Solid-phase extraction was used to extract PPX and IS from human plasma. The resolution of peaks was achieved with 0.01 m ammonium acetate buffer (pH 4.4):acetonitrile (30:70, v/v) on a Discovery CN column. The total chromatographic run time was 3.0 min and the elution of PPX and IS occurred at approximately 2.32 and 2.52, respectively. The MS/MS ion transitions monitored were 212.10 --> 153.10 for PPX and 180.20 --> 107.30 for IS. The method was proved to be accurate and precise at linearity range of 20-3540 pg/mL with a correlation coefficient (r) of > or =0.999. The intra- and inter-day precision and accuracy values found to be within the assay variability limits as per the FDA guidelines. The developed assay method was applied to a pharmacokinetic study in human volunteers following oral administration of 0.25 mg PPX tablet. |
| |
Keywords: | pramipexole human plasma validation LC‐MS/ MS pharmacokinetics |