摘 要: | In this paper,the usual pulse echo overlap method for ultrasonic wave velocity measurement isdeveloped to pulse echo overlap method with alternative triggering.The principle of the method liesin that,when the first ultrasonic echo train,being given rise from the first transmitting pulse,is dis-played on an oscilloscope,the oscilloscope is triggered normally and synchronized with the transmittingpulse.But for the next echo train to be displayed,which is given rise from the second transmittingpulse,it will be triggered after a further time delay than that of the first one,which is equal to a roundtrip time of an ultrasonic pulse in the sample.Then the whole procedure is repeated again and againand on the oscilloscope a pattern of overlapped echo trains can be observd.The value of the delaytime can be obtained directly from a digital counter.By changing the scanning rate of the oscilloscopeand adjusting its triggering delay,any part of the overlapped pattern can be observed and studied. Based on this princi
|