首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Peak Profile Analysis in High Field Asymmetric Wave Ion Mobility Spectrometry
Authors:Chi-lai Chen  De-yi Kong  Xiao-hua Wang  Huan-qin Wang  Feng Shuang  Tao Mei
Institution:1.Institute of Intelligent Machines, Chinese Academy of Sciences, State Key Laboratory of Transducer Technology, Hefei 230031, China;Department of Precision Machine and Precision Instrument, University of Science and Technology of China, Hefei 230027, Chin2.Institute of Intelligent Machines, Chinese Academy of Sciences, State Key Laboratory of Transducer Technology, Hefei 230031, China
Abstract:High field asymmetric wave ion mobility spectrometry (FAIMS) is a powerful tool to detect and characterize gas-phase ions, while the unsolvable partial differential equation of ions moving in ion drift tube poses a big challenge to FAIMS spectral peak analysis. In this work, a universal and effective model of FAIMS spectral peak profile has been proposed by introducing ion trajectory and loss height. With this model, the influence of the structure of ion drift tube, dispersion voltages, compensation voltages, and carrier gas flow rate on the FAIMS spectral peak characteristics like peak shape, full width at half maximum and peak height is analyzed and discussed. The results show that the influence of different factors on the FAIMS spectral peak profile can be qualitatively described by the model which agrees with the experimental data.
Keywords:High field asymmetric wave ion mobility spectrometry  Loss height  Ion tra-jectory  Spectral peak profile model  Parameter optimization
点击此处可从《化学物理学报(中文版)》浏览原始摘要信息
点击此处可从《化学物理学报(中文版)》下载免费的PDF全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号