首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Threonine-FX dosimeter for food irradiation
Authors:S R Nilekani  B L Gupta
Institution:

Radiation Safety Systems Division, BARC, Trombay, Mumbai-400 085, India

Abstract:A detailed study for the spectrophotometric readout method for L-threonine powder, CH3CH(OH)CH(NH2)COOH], was done. In this method, 400 mg unirradiated/irradiated L-threonine powder was dissolved in 10 ml of a solution which contains 3×10−4 mol dm−3 ferrous ammonium sulphate and 1.7×10−4 mol dm−3 xylenol orange (XO) in aerated aqueous 0.17 mol dm−3 sulphuric acid (FX). The peroxy radicals produced from irradiated threonine oxidize ferrous ions and XO forms a complex with ferric ions as well as controls the chain length of ferrous ion oxidation. The plot of absorbance at 556 nm against dose is linear in the dose range 20–400 Gy and doses down to about 1 Gy can be measured using 10-cm path cells. Response of the dosimeter is independent of irradiation temperature above 20. A dose of 50 Gy–10 kGy can be measured dissolving 50 mg threonine powder in 10 ml of a solution which contains 3×10−4 mol dm−3 ferrous ammonium sulphate and 1.3×10−4 mol dm−3 XO in aerated aqueous 0.06 mol dm−3 sulphuric acid (FX). The plot of absorbance at 552 nm against dose is non-linear. However dosimeter shows linear dose response up to 1000 Gy. Irradiated threonine powder is stable for about 3 months. The reproducibility of the method is better than ±2%. This dosimeter is very useful as transfer dosimeter for food irradiation programme.
Keywords:Threonine dosimeter  FX solution  Spectrophotometric readout  Food irradiation dosimetry
本文献已被 ScienceDirect 等数据库收录!