Determination of iron(II) and total iron in environmental water samples by flow injection analysis with column preconcentration of chelating resin functionalized with N-hydroxyethylethylenediamine ligands and chemiluminescence detection
Authors:Hirata S  Yoshihara H  Aihara M
Institution:

a Chugoku National Industrial Institute, 2-2-2 Hiro-suehiro Kure, Hiroshima 737-0197, Japan

b Kinki University, 1 Takaya-Umenobe, Higashi-Hiroshima, 739-2116, Japan

Abstract:A flow injection analysis (FIA) technique for the determination of Fe(II) and total-Fe in environmental water samples has been developed with a high sensitivity. The resin used for preconcentration of iron was the macroporous resin, Amberlite XAD-4 functionalized by N-hydroxyethylethylenediamine (HEED) groups. The technique employed was FIA by combination of on-line chelate resin preconcentration and chemiluminescence detection (CL), using brilliant sulfoflavine and hydrogen peroxide reagent solutions. The interference by coexisting Fe(III) could be eliminated by addition of 1×10−6 mol of deferrioxamine B solution. The detection limits of Fe(II) and total-Fe were 0.80 and 0.36 nmol l−1 for 5.6-ml seawater samples with a concentration of 2 nmol l−1. The relative standard deviations for both samples were less than ±4%. A typical analysis for Fe(II) can be performed in 7.5 min. The technique was ascertained by comparing the analytical value of total-Fe with the certified value of Fe in the reference standard seawater CASS-3.
Keywords:Flow injection analysis  Amberlite XAD-4-HEED column  Fe(II)  Environmental water samples