Radon determination by activated charcoal adsorption and liquid scintillation measurement
Authors:A C Canoba  F O López  A A Oliveira
Institution:(1) Environmental Studies Division, Nuclear Regulatory Authority, Avenida Del Libertador 8250, (1429) Buenos Aires, Argentina
Abstract:A passive diffusion method for the determination of radon concentration has been optimised and calibrated. The device consists of a scintillation vial containing activated charcoal, a diffusion barrier and a desiccant agent. The response to diverse atmospheric humidity and variable exposure intervals was studied. The result is a detector independent of atmospheric humidity till 7 days of exposure. The method was compared with electret detectors (US EPA) with very satisfactory results. The advantages of this method are its simplicity, low cost, low detection limit, the total automatization of the measurement and its total independence of humidity to measure in a wide range of radon concentrations.